INFICON Composer Gas Concentration Controller User Manual

Page 109

Advertising
background image

4 - 13

IP

N 07

4-

28

9L

Composer Operating Manual

Ú

. . . . . . . . . . . . . . . . . . . . . . . . . Illuminates when the concentration meets

the condition for “steady" and the STEADY
relay is closed, see

section 4.3.6.1,

Description and Purpose of SYSTEM I/O
Connector, on page 4-22

. You may press

the ZERO switch to alter the Reference
Zero when this indicator is illuminated.

CONCENTRATION TARGET . . . . Illuminates when the measured

concentration or bulk flow is within the
defined upper and lower relay limits of the
target concentration.

RELAY LIMIT UPPER. . . . . . . . . . Illuminates when the concentration or bulk

flow has exceeded the upper limit and the
@ CONCENTRATION relay is open.

RELAY LIMIT LOWER . . . . . . . . . Illuminates when the concentration or bulk

flow is less than the lower limit and the
@ CONCENTRATION relay is open.

CELL TEMPERATURE . . . . . . . . . Illuminates when the transducer’s

measurement chamber is at the value
programmed for
CELL TEMPERATURE
+0.1 °C / -0.2 °C.

CELL TEMPERATURE STEADY . Illuminates when the transducer’s

measurement chamber is at the value
programmed for CELL TEMPERATURE
and has been stable for 10 seconds.

TOTAL FLOW FIXED . . . . . . . . . . Illuminates when the “Control Dilution

Flow” configuration switch has been
activated, see

section 4.3.6.6, CFGR2, on

page 4-26

.

FUZZY . . . . . . . . . . . . . . . . . . . . . Illuminates when the fuzzy logic control

choice has been selected for the selected
film program.