4.3.5.2.1 Numerical Display and Units Indicators

-.8.8.8.8.8.8.8 . . . . . . . . . . . . . . . . Numerical display for measured

composition in the operate mode. Provides
all numerical indication for the various
parameter’s values in the program and
auxiliary display modes. Indicates the
various user diagnostic modes and error
conditions, see

section 6.5, User

Diagnostics, on page 6-7

, in conjunction

with WARNING indicator described in

section 4.3.5.2.5 on page 4-12

. The

allowable range and resolution of individual
parameters are described in

section 8.3.5,

Setup Process Parameters, on page 8-14

.

sccm . . . . . . . . . . . . . . . . . . . . . . Illuminates in the program mode when the

parameter value to be entered is a flow
rate. Illuminates in the auxiliary display
mode when the parameter to be displayed
is either the bubbler (B) or the dilution (D)
flow.

°C . . . . . . . . . . . . . . . . . . . . . . . . . Illuminates in the program mode when the

parameter value to be entered is a
temperature. Illuminates in the auxiliary
display mode when the parameter to be
displayed is cell temperature (T).

MOLE % . . . . . . . . . . . . . . . . . . . . Illuminates in the normal operate mode,

Illuminates in the program mode when the
parameter value to be entered is a
concentration (mole fraction percent).

% . . . . . . . . . . . . . . . . . . . . . . . . . . Normally part of the MOLE % message in

the operate mode. Illuminates in the
program mode when the parameter value
to be entered is a percent (RELAY LIMIT).

SEC . . . . . . . . . . . . . . . . . . . . . . . . Illuminates in the program mode when the

parameter value to be entered is a time
(DEAD TIME & TIME CONSTANT)

Hz . . . . . . . . . . . . . . . . . . . . . . . . . Illuminates in the auxiliary display mode

when the parameter value to be viewed in
the numerical display is the resonant
frequency (F).