4.3.6.6 CFGR2

The second eight contact binary encoded switch block that is used for activating
various instrument features. The state of these switches is only read when
power is first applied to the Controller. Changing the state of the switch while
the unit is powered on does not result in a change of the selected features. See

Table 4-5

.

Table 4-4 RS-232C Communication Baud Rates

SW#4

SW#3

SW#2

RS-232C

(Baud Rate)

0

0

0

4800

0

0

1

9600

0

1

0

19200

0

1

1

38400

1

0

0

57600

1

0

1

9600

1

1

0

9600

1

1

1

9600

Table 4-5 Configuration SW#2 Descriptions

SW

Function

Details

1

CONCENTRATION
OUTput (MSB)

See

section 4.3.6.8 on page 4-27

& also

SW#5

2

CONCENTRATION
OUTput

See

section 4.3.6.8 on page 4-27

& also

SW#5

3

CONCENTRATION
OUTput (LSB)

See

section 4.3.6.8 on page 4-27

& also

SW#5