6 cfgr2, Section 4.3.6.6, Section 4.3.6.6 on – INFICON Composer Gas Concentration Controller User Manual
Page 122: Sw7 of cf, Tivated, see, Section 4.3.6.6, cfgr2, on, Table 4-4
Advertising
4 - 26
IP
N 07
4-
28
9L
Composer Operating Manual
4.3.6.6 CFGR2
The second eight contact binary encoded switch block that is used for activating
various instrument features. The state of these switches is only read when
power is first applied to the Controller. Changing the state of the switch while
the unit is powered on does not result in a change of the selected features. See
.
Table 4-4 RS-232C Communication Baud Rates
SW#4
SW#3
SW#2
RS-232C
(Baud Rate)
0
0
0
4800
0
0
1
9600
0
1
0
19200
0
1
1
38400
1
0
0
57600
1
0
1
9600
1
1
0
9600
1
1
1
9600
Table 4-5 Configuration SW#2 Descriptions
SW
Function
Details
1
CONCENTRATION
OUTput (MSB)
See
& also
SW#5
2
CONCENTRATION
OUTput
See
& also
SW#5
3
CONCENTRATION
OUTput (LSB)
See
& also
SW#5