3 power connector, 4 option connector, 5 cfgr1 – INFICON Composer Gas Concentration Controller User Manual

Page 121: Section, Section 4.3.6.5

Advertising
background image

4 - 25

IP

N 07

4-

28

9L

Composer Operating Manual

4.3.6.3 POWER Connector

A DB-9 connector used for connection to a 24 volt power supply. Refer to

section 1.7 on page 1-7

for a full description of the power supply requirements

and pin connections.

4.3.6.4 Option Connector

A DB-9 connector used for:

Š

Bulk Flow Control configuration: for providing Concentration read back
output and MFC read-back input.

Š

Concentration Control configuration: to provide concentration output for a
Composer using dilution flow in the concentration control mode.

4.3.6.5 CFGR1

The first of 2 eight contact binary switch blocks used for activating various
features. The state of these switches is only read when power is first applied to
the Controller. See

Table 4-3

for the configuration switch number descriptions

and

Table 4-4

for the baud rates.

NOTE: Changing the state of a switch while the Controller is powered on does

not result in a change of the selected features. Switch status is only
read during the Controller’s power on sequence.

Table 4-3 Configuration Switch #1 Descriptions

SW#

FUNCTION

Details

1

transducer type
refer to

section 1.10.5.5 on page 1-23

0 = off - series 19 and higher
1 = on - series 18 and lower

2

communications configuration

(see note below)

3

communications configuration

(see note below)

4

communications configuration

(see note below)

5

spare

6

always use factory set Reference Zero (0=off, 1=on) When on,

Reference Zero may not be
user set.

7

parameter lock

(0=off, 1=on) When on,
parameters may not be altered
from the front panel. However,
all parameters may be changed
through communications.

8

spare

NOTE: The supported communication is RS-232C.