6 error detection, 1 led blink patterns, 2 error codes – INFICON Composer Gas Concentration Controller User Manual

Page 142

Advertising
background image

6 - 10

IP

N 07

4-

28

9L

Composer Operating Manual

6.6 Error Detection

Specific errors can be detected in one of the following ways:

1

Observing the blink pattern on the Warning LED, refer to

Table 6-7 on page

6-7

for a complete list of error or warning indicated.

2

On the optional LCD Display. When an error occurs, an error code will
replace the concentration output. If there is a warning (less serious
problem) the "Warning" annunciator will be displayed; by pressing key 5 a
warning code will be displayed. For example, if there is no gas in the
transducer the display will show "ER 80", because no measurable signal
was found. A complete listing of errors/warnings is given in

Table 6-7 on

page 6-7

.

3

Via RS-232C; read the error log.

4

Via the Monitor Software pop up window.

6.6.1 LED Blink Patterns

Refer to

Table 6-7 on page 6-7

.

6.6.2 Error Codes

On the LCD and in the error log, errors are uniquely coded in hexadecimal with
each bit representing a different error. This way, multiple errors can be
recorded. Refer to

Table 6-7 on page 6-7

for error codes. Multiple errors are

decoded as follows:

Example 1:
There are two errors present.

The displayed error is 0x0084
It is caused by:

Ambiguous Gas Warning

0x0004

No Signal Error

0x0080

Error Code summed

0x0084

Example 2:

There are three errors present.
The displayed error is 0x001C
It is caused by:

Identical Gasses

0x0010

Bad Peak Shape

0x0008

Ambiguous Gasses

0x0004

Error code summed

0x001C