1.8 Environmental Requirements
The following paragraphs detail the Composer’s environmental requirements.
Use
This instrument is designed for indoor use only.
Altitude range
The Composer can be used to a maximum altitude of 2000 meters
(6561 feet). For applications above this altitude, please consult the factory.
Maximum humidity
Non-condensing relative humidity at operating temperature.
Pollution degree
Category II (as defined by EN61010-1, only non-conductive
pollution occurs).
Operating temperature
The maximum and minimum operating temperature is shown in the following
sections.
Maximum temperature
Controller: 50 °C
Transducer: 65 °C
Minimum temperatures
Controller: 0 °C
Transducer: 0.1 °C (minimum); recommended operating is 5 °C above
ambient for maximum stability.
1.9 Computer System Requirements
The Composer utilizes the Setup Tool software or the Monitor software for
monitoring and storing data. See
for the computer
system requirements.
1.10 Installation
The Composer should be installed following the guidelines presented in the
following sections.