Instrument setup – Leica Biosystems IPS User Manual

Page 25

Advertising
background image

23

Leica IP S

• Unscrew the 8 screws (2,

Fig. 8

) on the sides

of the wooden box and loosen the cover.

• Remove the accessory box (7,

Fig. 9

) including

accessories and packaging material) - directly

under the lid.

• Unscrew the 8 screws (3,

Fig. 8

) on the floor

of the wooden box on the exterior.

• Remove the interior carton around the instru-

ment.

• Carefully remove the wooden box from the

baseplate.

• The printer is secured with 4 plates (5) on

the wooden floor of the outside box. Loosen

the two screws (4,

Fig. 10

) on the base of the

instrument. Remove the plate from the bottom.

• Place the printer from the baseplate on a

stable laboratory bench - or, if present, on the

automated unload station. Make sure that the

stage is leveled!

Fig. 9

7

4

5

5

Fig. 10

When unpacking the printer, at least

two people (one person on each side

of the printer) are required to lift the

printer out of the box and place it onto

the laboratory bench.

Fig. 11

6

• If the instrument was set up at its final area of

use, remove the blue foam transport anchor

(6,

Fig. 11

) (pull upwards).