5. Operation

Warning:

Problem with cassette ejection from a magazine

81 The display consists of two parts: The "8" is a warning that a magazine

ejector is mechanically blocked. The second digit of the

83 message gives the number of the affected magazine.

Error messages
All displayed numbers between 20 and 78 and 89 and 93.

Display

14 Prompt requesting maintenance

If this message is displayed, the instrument will be due for main-

tenance within the next few weeks. Confirm the prompt pressing

ERROR.

After about 3 weeks the message will be displayed again and will not

disappear from the screen when pressing ERROR.

You can still continue printing but maintenance must be carried out

urgently.

15 Cleaning the print head

Screen prompt requesting the operator to manually clean the print

head (cleaning swab + alcohol).

The printer is off-line. No new print jobs are accepted.

87 After the last cartridge change, CLEAN has been pressed to indi-

cate to the printer that a cleaning cartridge has been inserted.

The printer has received a print job but is unable to print because

the cartridge contains cleaning fluid instead of ink.

Remedy:

Cancel the print job. Switch the printer off and back on and change

the cartridge. Then press LOADED or ERROR and wait for 2 minutes.

88 Ink cartridge has been changed, device waits for confirmation with

ERROR, CLEAN, or LOADED key.

Status messages (continued)

Caution!

NEVER press LOADED after reinserting an ink cartridge which has

already been used. This could cause permanent damage to the

printer.