6. Cleaning and Maintenance

Outer surfaces
• Clean the outer surfaces (including those of

the automated cassette unload station) with

a mild detergent and subsequently dry with a

slightly moistened cloth.

• Do not use any solvents for cleaning the outer

surfaces and the lid!

Automated unload station
• Remove the unload trays; with a brush, remove

dust and debris from guides and ejector.

• The trays themselves can be cleaned with a

household cleaner.

• Do not use any solvents to clean the trays!
• Prior to reinserting them into the instrument,

the trays must be completely dry.

• Drying station (

Fig. 51

)

The chute must be clean.

Fig. 51

Slide guiding mechanisms

• The drawer for broken glass (77) is located

above the slot for the ink cartridge, behind

the left cover (5). It prevents glass dust and

broken glass from falling into the interior of

the instrument. The drawer can be pulled out

sideways by grasping the black handle (78)

and must be cleaned regularly (

Fig. 52

).

• Broken glass can easily be removed - use a

brush and remove any glass through the open-

ing in the center of the drawer.

Fig. 52

77 78

5

Drawer for broken glass

Caution!

Sensitive electronics components are

located in this area.

Use no liquid in this area!