7. Troubleshooting

Error messages

(continued)

Display

code

Source of error

Troubleshooting solution

Cover (

4 in Fig. 2

) not closed properly.

Safety switch triggered.

Ensure that the cover is not blocked,

e.g. by the manual unload station.

Close cover completely.

Flash standby state not reached within the prescribed

time. Charging electronics defective.

Call Leica Technical Service.

No flash or flash duration too short.

Flash bulb dirty or defective.

Check whether flash bulb works – to do

so, observe the stray light on the cover.

Never open the cover to check wheth-

er a flash is triggered!

No flash --> install a new flash bulb.

See

Chapter 4.7

.

The temperature in the drying module is not within the

permissible temperature range.

Carry out a RESET. See

Chapter 7.1

.

Unload station mechanically blocked.

Remove the cause of the blockage.

Heater fan not running or running too slowly.

Call Leica Technical Service.

No slide in the drying module for flashing. Slide left the

specimen slide carrier, but did not reach the drying

module.

Remove the specimen slide.

Slide not successfully ejected from drying module.

Remove the specimen slide.

Slide in the drying module during initialization or prior to

a print job being carried out.

Remove the cause of the blockage.

The flap of the drying module either does not close or

does not open; flap is blocked (e.g. by a slide).

Remove the cause of the blockage.

Lifting table end sensor does not switch.

Call Leica Technical Service.

Lifting table position sensor does not switch during

positioning.

Call Leica Technical Service.

Flash power supply: Overtemperature for more than

10 minutes.

Call Leica Technical Service.