Instructions for Use V 1.9 RevD - 05/2013

5. Operation

"ADVANCED OPTIONS" menu
Clicking onto the individual menu items opens up

a pull-down menu to their right, where you can

select the desired parameters.

Any menu items not described here are of no im-

portance for the printer. Therefore, the standard

settings of all menu items not described here

should remain unchanged.

• In the PAPER SIZE menu you select the size of

the imprintable zone of the

slide. Slide must be selected

in this menu, otherwise

printing will not be possible.

• Selectable print head RESOLUTION settings

are 360 and 180 dpi. With

slide surfaces not appropri-

ate for 360 dpi resolution,

printing results will be poor.

For such cassettes, 180 dpi should be selected.

• The PAPER/OUTPUT menu item is important

above all for the multi-cas-

sette unload station.

SAME TRAY:

slides keep being deposited into a tray until

the tray is full.

JOB IN NEW TRAY:

each print job starts with a new tray.

NEW TRAY:

Only for special applications—do not select

this option under standard Windows programs.

Fig. 48

PRINT QUALITY menu item

In this menu, you can select

whether an imprint is to be

applied onto a specimen

slide once (NORMAL) or

twice (OVERSTRIKE).

UPSIDE DOWN imprints are rotated 180°. Here

again, OVERSTRIKE can be selected.

Due to the higher ink consumption and

the substantially poorer printed image,

the OVERSTRIKE setting may not be

used.

When using the manual unloading

system, the values configured in the

PAPER/OUTPUT menu item are not

considered by the device.