Instrument setup – Leica Biosystems IPS User Manual

Page 37

Advertising
background image

35

Leica IP S

4. Instrument Setup

Fig. 40

87

86

• Remove the ink cartridge (66) from the delivery

carton and shake it a few times, then remove

the protective foil.

• The ink cartridge must be replaced after

60,000 printouts or 3.5 months. Record the

installation date on the white surface (65) on

the front (

Fig. 39

).

• Insert the cartridge into the slot. Loosen the

red screw-on cap one full turn but do not

remove it yet.

• Insert the cartridge completely into the recep-

tacle applying moderate pressure. You have

to use a certain amount of force to pierce the

seals.

Withdraw the cartridge by about 1� inch

(30 mm) and push it back in as far as it will

go; repeat this step two or three times.

• Unscrew the red cap (87) completely from the

nozzle of the cartridge and store the screw-on

cap in the groove indented into the cartridge

for that purpose (

Fig. 38, 39

).

Exchanging the cartridge

(continued)

Fig. 37

Fig. 39

Fig. 38

65

66

Very important!

Prior to each transport, the cap (87)

must be screwed onto the nozzle to

prevent the ink from spilling.

This step is very important to ensure

safe operation of the printer!

• Directly next to the slot opening is a safety

catch (86,

Fig. 40

). Moving this catch to the

side secures the ink cartridge against falling

out.