26

Instructions for Use V 1.9 RevD - 05/2013

4. Instrument Setup

4.5 Installing the manual unload station

Fig. 13

27

26

28

29

32

The enlarged detail shows

the correct location of the

collar screw once the unload

station has been locked in

place correctly.

30

31

31

4

28

The unload station supplied consists of:
26 - Unload station

27 - Screening plate

28 - Collar screws (3 pcs.)

29 - Slotted screws with washers (2 pcs.)

Install as follows (

Fig. 13

):

• Open lid (4).

• There are 5 tapped holes (2 x item no. 30 and

3 x item no. 31) in the installation surface

located below the reflector.

• With a screwdriver, insert 3 collar screws (28)

in tapped holes (31) as far as they will go.

• Then, fasten cover (27) in tapped holes (30)

using the two slotted screws (and washers)

(29).

• To fasten the unload station to the instrument,

place the wider end of the three oblong holes

(32) over the heads of the three collar screws

(28). Press the unload station against the in-

stallation surface, pushing it simultaneously to

the right until it locks in place (see enlarged

detail).

If the unload station does not easily slide past

the shielding cover, slightly lift the front end of

the device.

• Close lid, making sure the unload station does

not obstruct the lid.

When working with the manual un-

load station the slides imprinted must

be removed in regular intervals. Oth-

erwise they will start piling up in the

unload station, printing will stop and

error code "44" will be displayed.