4. Instrument Setup

4.10 Exchanging the cartridge

Fig. 31

4

5

64

85

The printer is supplied with a cleaning cartridge

(64) inserted. To be able to print, the cleaning

cartridge must be exchanged for an ink cartridge.
To do so, proceed as follows:
• Open the cover plate (5) on the left side of the

instrument (pressing its top left corner).

• Unscrew the red cap (85) of the cleaning car-

tridge (64) by one turn (

Fig. 31

), then switch

on the printer using the main switch on the

rear side (item (62) in

Fig. 23

) and wait for it to

initialize.

• Open the hood (4 in

Fig. 31

), then

press the CLEAN and LOADED

keys simultaneously.

• The print head (72) moves

upward to a position approx.

1 cm away from the sealing lip

(

Fig. 32

).

• Raise the lever (69) then remove the black

transport plate (73) and insert the location

plate (70), which is required for printing

(

Fig. 33

).

Fig. 33

Fig. 32

72

73

69

73

70

Do not reinstall a used transport plate

(73), as it will no longer completely

seal the print head.

To prevent damage to the print head,

always use the red location plate (70)

when printing.