Leica IP S

4. Instrument Setup

• The contact spring (44) has to touch the igni-

tion wire (45) of the lamp after being inserted.

• Swing the reflector downwards. Reinsert and

retighten screw (49).

• Close lid (4) of the instrument again.

Fig. 20

45

47

Inserting the new flash bulb

47

48

Fig. 22

Fig. 19

44

46

Fig. 21

45

44

85

86

86

Caution!

If the lamp electrodes are inserted the

wrong way, the flash bulb will still

work, but the lifetime will be signifi-

cantly reduced.

• Insert the shielding glass (85) in the two hold-

ers (86).

• Insert the new flash bulb into socket (46)

(

Fig. 21

); then push it carefully inwards as far

as it will go (

Fig. 22

) (the polarity marker (+)

must not be visible any longer). If necessary,

move flash bulb gently up and down.

• Important! Make sure that the new flash bulb

is inserted correctly – the lamp electrode

(47) marked with a + must be inserted into

the socket (46) that bears the same mark (48)

(

Fig. 21

).