7. Troubleshooting

Malfunction / failure
• Instrument not functioning.

• No display indication.

• Instrument not working at normal speed.

Printing a slide takes approx. 8 sec., even

after the warm-up phase has been com-

pleted.

• Check whether there is a general power failure (no power).
• Check whether the power plug is inserted correctly into the wall outlet

and whether the wall outlet is switched on, if applicable.

• Check whether the power switch is switched on correctly. Possibly, the

primary fuse has tripped. If so, the main switch will be in "0" = OFF posi-

tion (

Fig. 63

).

• Check whether one of the two secondary fuses is defective. (F1, F2

located at the back panel of instrument,

Fig. 64, see also Fig. 65

).

Some instrument malfunctions / failures are caused by defective fuses.

Fuse F1

Fuse F2

7.5 Power failure

Fuse to be checked

7.4 Changing the flash bulb

Code 13 is displayed when the flash bulb has reached its maximum service-

able life.
When code 13 is issued, the bulb has to be changed.
For details on how to insert / replace the flash bulb, please refer to

Chap-

ter 4.7

.

Fig. 64

Fig. 63

10 s

Before replacing the flash bulb, switch off the printer and unplug

the instrument from power supply.

After changing the bulb, switch the printer back on.
Then go to Off-line mode and press and hold the ERROR button for 10 sec-

onds - code 13 disappears from the display.