2 status messages, Status messages, Troubleshooting – Leica Biosystems IPS User Manual

Page 62

Advertising
background image

60

Instructions for Use V 1.9 RevD - 05/2013

7.2 Status messages

Display

code

LED

Meaning

(For additional information, refer also to

Chap. 5.2

)

7. Troubleshooting

"Mag. empty flashing"

The printer is waiting for an individual cassette to be placed in the

feeding chute for printing.

"Mag. empty flashing"

Magazine No. 1 empty

"Mag. empty flashing"

Magazine No. 2 empty

"Mag. empty flashing"

Magazine No. 3 empty

–

Ink print head cleaning is in progress.

–

Temperature in the flash power supply is too high.

–

Drying module temperature too low / too high.

–

Flash bulb has reached its maximum life. (See also

Chapter 7.4

)

–

Prompt requesting maintenance.

–

Prompt to carry out a head cleaning.

"Ink empty" flashing

Ink cartridge has been changed; instrument waiting for confirma-

tion via ERROR, CLEAN, or LOADED button.