7 installing/exchanging the flash bulb, Installing/exchanging the flash bulb, Instrument setup – Leica Biosystems IPS User Manual

Page 30

Advertising
background image

28

Instructions for Use V 1.9 RevD - 05/2013

4. Instrument Setup

Removing the old flash bulb

4.7 Installing/exchanging the flash bulb

49

41

Fig. 15

• Swing the reflector (41) upwards.

• Carefully pull out the old flash bulb (43) straight

to the right, do not twist it. If the flash bulb

cannot be pulled out easily, gently rock it back

and forth to get it unseated from the socket.

• Make sure the contact spring (44) is removed

from the priming wire (45) of the lamp.

(See also

Fig. 19

and

22

)

Fig. 16

Fig. 17

• Open the lid (4,

Fig. 2

) to gain access to the

reflector.

• Remove screw (49) (use screwdriver supplied

as part of tool set). Mind the washer (42).

Fig. 18

42

43

44

Switch the instrument off and unplug it

from power supply.

Allow the flash tube to cool before re-

moving it. Do not handle the flash tube

with bare hands. Use a glove or tissue.

Caution!
To insert/remove, hold the flash

bulb as shown in

Fig. 17

(right).

Do not take hold of the sides of

the flash bulb or compress it as

shown in

Fig. 16

.