Instrument setup – Leica Biosystems IPS User Manual

Page 41

Advertising
background image

39

Leica IP S

Configuring the serial port:

1. Open the printer folder:

START --> SETTINGS --> PRINTERS
2.
If you right-click on the icon of the newly installed printer, a quick menu

opens up.

Click on PROPERTIES....
3. The PROPERTIES window contains several tabs; select the PORTS tab.

4. The port selected during installation (COM 2 in the example here) must

be highlighted. Enable the button ENABLE BIDIRECTIONAL SUPPORT.

5. Click on CONFIGURE PORT... to open the Ports dialog box. Select the

corresponding port and click on SETTINGS.... The SETTINGS FOR... dia-

log box opens up. Set BAUD RATE to 57600. Leave all other settings as

shown in

Fig. 41

.

Caution!

Windows 2000 and XP do not

contain the PORTS page.

Fig. 41

4. Instrument Setup