2 clean print head, Clean print head, Cleaning and maintenance – Leica Biosystems IPS User Manual

Page 55

Advertising
background image

53

Leica IP S

6. Cleaning and Maintenance

Once a week or if message "15" is

displayed, the print head must be

cleaned manually.

Preparing the printer

• Open the printer lid and then

press buttons CLEAN and

LOADED simultaneously.

• The print head (72) moves

upward to a position approx.

1 cm away from the sealing lip

(

Fig. 53

).

6.2 Clean print head

• Carefully insert the swab (89) into the gap

under the print head and apply light pres-

sure upwards (on print head) and move the

foam swab back and forth (

Fig. 55

) (approx.

10 times!).

This procedure removes dried ink residues.

Fig. 53

69

70

+

• Push the lever (69) upwards, then remove

the red location plate (70) with the sealing lip

(

Fig. 54

).

• Moisten one of the foam swabs supplied with

the instrument (73) with some alcohol. Be sure

not to use too much alcohol—no alcohol may

drip into the instrument.

Fig. 54

Fig. 55

72

89

Never use acetone or xylene! Only use

alcohol 95 % or 100 % for cleaning pur-

poses.

Never rotate the swab—this can dam-

age the nozzle plate of the print head.

Advertising