3 recalculation rules – Metrohm viva 1.1 User Manual

Page 297

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Database

viva 1.1

■■■■■■■■

285

This button is disabled for as long as recalculation has not yet been trig-
gered and if not all the selected determinations were able to be recalcu-
lated.

[Cancel] or

Closes the Reprocess dialog window. The result of the reprocessing
made since the last saving will not be saved.

Reprocessing with history
Versioning behavior of reprocessed determinations.

Selection

On | Off

Default value

Off

On
A new version is created and saved each time a determination is repro-
cessed.
Off
The version number is increased with each reprocessing of a determi-
nation, but only two versions are stored in the database (the first, origi-
nal version and the latest, reprocessed version). In the case of already
reprocessed determinations, the version that was last saved is deleted
and a new version is created.

4.5.2.5.3

Recalculation rules

Dialog window: Reprocess

The following rules apply for recalculating the selected and modified
determinations:

■

Number of determinations
No more than 10 determinations may be selected for recalculation.

■

Start test
During the start test all device tests and device monitoring will be
ignored.

■

Special tracks
Series start track, series end track and error tracks are not run through
during recalculation.

■

Tracks with "Return immediately"
If the method contains tracks with the option Return immediately
enabled, then a warning will will appear, because it can no longer be
guaranteed that during recalculation the method will run in the same
way as during the determination.