6 Configuration

viva 1.1

■■■■■■■■

805

Concentration unit

Concentration unit for the lower or upper limit value of the calibration
range.

Selection

mol/L | mmol/L | µmol/mL | g/L | mg/L | µmol/L |
mg/mL | µg/mL | ppm | % | mEq/L

Default value

ppm

Calibration temperature

Temperature at which the calibration is carried out. In addition, the display
shows whether the temperature was measured with a Pt1000 or an NTC
temperature sensor or entered manually.

Input range

-20.0 - 150.0 °C

Default value

25.0 °C

NOTE

If the Calibration temperature has been entered manually, then the
following parameters will be reset to the default value or updated:

Calibration temperature = manual

Calibration date is updated.

Calibration method = manual

Determination ID is deleted

User is updated

Calibration date

Shows the date and time of the last calibration entered after each auto-
matic calibration or manual entry.

Calibration method

Display of the method name with which the last calibration has been car-
ried out. If the calibration data has been entered manually, then manual
will appear.

Determination ID

Identification of the determination.

The determination ID can be copied and pasted into another field.

User

Shows the short name of the user who was logged in during the calibra-
tion or who entered the calibration data manually. If the login function is
not used, then the user logged in under Windows will be entered auto-
matically.