Metrohm viva 1.1 User Manual

Page 643

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1

■■■■■■■■

631

Result variables if called by CALL VA
The following result variables are generated in the method run when a
CPVS command is called by CALL VA and can be used in formulas:

Result variables per substance, variation and replication

Designation 'RS.'Command name'.'Substance
name'.VAR{x}.REP{y}.Variable identification'

Identification

Description

SUBST

Substance name

POT

Stripping potential

AREA

Peak area

HGT

Peak height

START

Start base point, where the peak evaluation baseline
begins

END

End base point, where the peak evaluation baseline
ends

AREANORM

Standardized peak area (AREA/AREA.MNV).

AREA.MNV = VMS.AREA.MNV

1

or ELECTRO-

LYTE.AREA.MNV

2

(only one of these two variables

may be present in a determination)

HGTNORM

Standardized peak height (HGT/HGT.MNV).

HGT.MNV = VMS.HGT.MNV

1

or ELECTRO-

LYTE.HGT.MNV

2

(only one of these two variables

may be present in a determination)

1

for calibration method DT

2

for calibration method RC

Result variables per substance and variation

Designation 'RS.'Command name'.'Substance name'.VAR{x}.Varia-
ble identification'

Identification

Description

AREA.MNV

Mean value of the peak areas of all replications

AREA.ASD

Absolute standard deviation of the peak areas of all
replications

AREA.RSD

Relative standard deviation of the peak areas of all
replications

Advertising