8 Manual control

viva 1.1

■■■■■■■■

963

[Fill]

Starts filling of the dosing cylinder for the selected dosing device. This but-
ton is only present if Fill automatically is switched off. The volume dis-
play shows the added volume until the filling procedure has ended.

[Stop]

Stops fixed volume dosing for the selected dosing device. If dosing has
been stopped, it cannot be resumed.

8.4.2.7

Dosing - Dosing

Program part: Manual control

Manual dosing can be started and stopped here. This tab is only displayed
if a single dosing unit is selected.

Dosing rate

Rate at which dosing should take place. The maximum dosing rate
depends on the cylinder volume of the dosing unit used. If the entered
dosing rate is too high for the selected dosing device, it will automatically
be reduced during dosing to the largest possible value.

NOTE

The dosing rate should be reduced for viscous liquids.

Selection

maximum

Default value

maximum

Input range

0.01 - 166.00 mL

Default value

4.00 mL

Filling rate

The dosing cylinder will be filled at this rate after dosing. The maximum
filling rate depends on the cylinder volume of the dosing unit used. If the
entered filling rate is too high for the selected dosing device, it will auto-
matically be reduced during filling to the largest possible value.

NOTE

The filling rate should be reduced for viscous liquids.

Selection

maximum

Default value

maximum

Input range

0.01 - 166.00 mL

Default value

4.00 mL