4 meas spec - evaluations, 5 meas spec - evaluation window – Metrohm viva 1.1 User Manual

Page 528

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

516

■■■■■■■■

viva 1.1

Input range

0.01 - 600,000 ms

Default value

6 ms

Averaged spectra

The number of spectra that are acquired and averaged.

Input range

1 - 10,000

Default value

10

Smoothing

Number of neighboring pixels with which the value for each measurement
pixel is determined.

Input range

0 - 100 pixels

Default value

0 pixels

5.5.2.7.3.4

MEAS Spec - Evaluations

Tab: Method

▶ MEAS Spec ▶ Properties... ▶ Evaluations

Command name

Name of the command.

Entry

25 characters

The evaluation windows for the evaluation of measurement curves can be
activated and defined on this tab.

[New]

Opens the Evaluation window # dialog window for entering the start
and end wavelengths (see Chapter 5.5.2.7.3.5, page 516).

[Properties]

Opens the Evaluation window # dialog window for editing the start
and end wavelengths selected in the table (see Chapter 5.5.2.7.3.5, page
516)
.

[Delete]

Deletes the selected line.

5.5.2.7.3.5

MEAS Spec - Evaluation window

Dialog window: Method

▶ MEAS Spec ▶ Properties... ▶ Evaluations ▶ [New] /

[Properties]

▶ Evaluation window #

Start wavelength

Lower limit of the spectrum.

Input range

100.0 - 2,000.0 nm

Default value

400.0 nm