Metrohm viva 1.1 User Manual

Page 463

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1

■■■■■■■■

451

Only editable for Addition = Add manually, otherwise disabled (includ-
ing text input field).

Selection

Display standard message | Display message
defined by the user

Default value

Display standard message

Display standard message
If this check box is activated, then the standard message for additions
will be displayed.
Display message defined by the user
If this check box is activated, then the message defined in the text field
will be displayed.

Dosing device
Parameters for automatic dosing. This section is displayed only when
Addition = Add with dosing device.

Dosing rate

Rate (volume per time unit) of expulsion. The maximum rate depends on
the cylinder volume of the dosing unit used and on the diameter of the
tubing at the selected dosing port.

Input range

0.01 - 166.00 mL/min

Default value

2.00 mL/min

Selection

maximum

CAUTION

The maximum permitted dosing rate of 4 mL/min may not be exceeded
for capillaries with a diameter < 1 mm.

Filling rate

Rate (volume per time unit) of filling. The maximum rate depends on the
cylinder volume of the dosing unit used and on the diameter of the tubing
at the selected fill port.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum