5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

534

■■■■■■■■

viva 1.1

5.5.2.7.5.3

MEAS Opt Conc - Measuring parameters

Tab: Method

MEAS Opt Conc Properties... MEAS Opt Conc - 'Command

name'

Measuring parameters

Command name

Name of the command.

Entry

25 characters

The parameters for the measuring procedure can be defined on this tab.

Tab for Avantes spectrometer

Wavelength

Wavelength at which the measured quantity (absorbance) is recorded.

Input range

100.0 - 2,000.0 nm

Default value

400.0 nm

Blank value

Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.

Input range

–1E+12 - 1E+12 (max. 10 digits)

Default value

0.0

Entry

1 - 10 characters

Selection

mol/L | mmol/L | µmol/L | g/L | mg/L | µg/L |
mg/mL | µg/mL | ppm | % | mEq/L

Default value

ppm

Integration time

Integration time for absorbance measurement.

Input range

0.01 - 600,000 ms

Default value

6 ms

Averaged spectra

Number of spectra that are recorded and averaged.

Input range

1 - 10,000

Default value

10

Smoothing

Number of neighboring pixels with which the light intensity of each mea-
surement pixel is averaged.