Page 499

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1

■■■■■■■■

487

Sensor

Colorimetric sensor

Selection of a colorimetric sensor from the table of colorimetric sensors.
The calibration data for the colorimetric sensor will be adopted for the
determination.

Selection

'Name of the colorimetric sensor' | empty or
'first list entry'

Default value

empty or 'first list entry'

Temperature measurement

Type of temperature measurement

Displayed only when Device type = 089 Photometer.

Selection

continuous | off

Default value

continuous

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Measuring parameters tab will be
used.

Stirrer
This section is displayed only when Device type = 089 Photometer.

Stirrer

Selection of the stirrer.

Selection

1 | off

Default value

1

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate.

Input range

0 - 15

Default value

3