5 Method

viva 1.1

■■■■■■■■

485

Command variables
The following command variables are generated by the CAL MEAS Opt
command in the method run and can be used in formulas under the des-
ignation 'Command name.Variable identification':

Identification

Description

BSY

Command status:

1 = BUSY, HOLD or ERROR

0 = READY

invalid (variable not available) = Command has
never been started

DBL

Total duration for the processing of the command in
s

EME

End measured value (measured value after process-
ing of the command) in the unit of the measured
value

ETE

1)

End temperature (temperature after the command
has been processed) in °C

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

invalid (variable not available) = Command has
never been started

IME

Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value

ITE

1)

Initial temperature (temperature before start condi-
tions are processed) in °C

LP.CNT

Transmission of the last measuring point in the mea-
suring point list in %

LP.MEA

Measured value for the last measuring point in the
measuring point list in the unit of the measured
value

LP.TEM

Temperature for the last measuring point in the mea-
suring point list in °C

LP.TIM

Time in s until the last measuring point in the mea-
suring point list is reached