5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

452

■■■■■■■■

viva 1.1 (for process analysis)

5.5.2.4.3

ADD STD

5.5.2.4.3.1

ADD STD - Overview

Dialog window: Method

ADD STD Properties... ADD STD - 'Command name'

Command for addition of standard solutions.

Appearance
The command has the following appearance:

Parameters
The parameters for the ADD STD command are configured in the follow-
ing dialog window:

ADD STD - Properties

Command variables
The following command variables are generated by the ADD STD com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':

Identification

Description

ACO

Addition counter (Number) = current number of
completed runs of the command

BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

CYL

Cylinder volume of the dosing unit used for the com-
mand

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

not valid (variable not available) = Command
has not yet been started

VOL

Dosed volume increment in L (Number)

VOLTOT

Dosed total volume in L (Number)