Metrohm viva 1.1 (ProLab) User Manual

Page 945

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Devices

viva 1.1 (for process analysis)

■■■■■■■■

933

of the control value that a PD controller causes immediately as a result of
the D component.

for Control task = Pump control
Input range

0 - 9,999.0

Default value

0.0

for Control task = Free controller
Input range

0.0 - 9,999.0 s

Default value

0.0 s

Damping time

for Control task = Pump control
Input range

0.0 - 9,999.0

Default value

0.0

for Control task = Free controller
Input range

0.0 - 9,999.0 s

Default value

0.0 s

D in feedback path

on | off (Default value: off)

Upper limit

Upper limit value at which the integration is stopped and the output is
limited.

Input range

0% mapping of the output - 100% mapping
of the output 'Mapping unit of the output'

Default value

100% mapping of the output 'Mapping unit
of the output'

Lower limit

Lower limit value at which the integration is stopped and the output is
limited.

Input range

0% mapping of the output - 100% mapping
of the output 'Mapping unit of the output'

Default value

0% mapping of the output 'Mapping unit of
the output'

ARW on I part only

on | off (Default value: off)