7 Devices
viva 1.1 (for process analysis)
■■■■■■■■
933
of the control value that a PD controller causes immediately as a result of
the D component.
for Control task = Pump control
Input range
0 - 9,999.0
Default value
0.0
for Control task = Free controller
Input range
0.0 - 9,999.0 s
Default value
0.0 s
Damping time
for Control task = Pump control
Input range
0.0 - 9,999.0
Default value
0.0
for Control task = Free controller
Input range
0.0 - 9,999.0 s
Default value
0.0 s
D in feedback path
on | off (Default value: off)
Upper limit
Upper limit value at which the integration is stopped and the output is
limited.
Input range
0% mapping of the output - 100% mapping
of the output 'Mapping unit of the output'
Default value
100% mapping of the output 'Mapping unit
of the output'
Lower limit
Lower limit value at which the integration is stopped and the output is
limited.
Input range
0% mapping of the output - 100% mapping
of the output 'Mapping unit of the output'
Default value
0% mapping of the output 'Mapping unit of
the output'
ARW on I part only
on | off (Default value: off)