2 port - properties – Metrohm viva 1.1 (ProLab) User Manual
Page 587
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
575
5.5.2.8.8.2
PORT - Properties
Dialog window: Method
▶ PORT ▶ Properties... ▶ PORT - 'Command name'
Command name
Name of the command.
Entry
25 characters
Device
Device name
Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Shows the instrument type.
Selection
846 Dosing Interface
Default value
846 Dosing Interface
Multiport valve
Multiport valve
Select the MSB connector the multi-port-valve is connected to.
Selection
1 | 2 | 3 | 4
Default value
1
Port
Selection of the target port to be switched to.
Input range
1 - 12
Default value
1
Rotating direction
Selection of the direction in which the valve moves to the target port.
Selection
ascending | descending | automatic | not over
Default value
automatic
ascending
The valve moves towards the target port in ascending direction.