5 Method

viva 1.0 (for Process analysis)

■■■■■■■■

589

5.6.5

Evaluation - Calibration

5.6.5.1

Calibration- Overview

Subwindow section: Method

▶ Evaluation - Calibration ▶

Application
The way that the sample content for the substances is determined from
the peak data is defined in the Evaluation - Calibration section of the
evaluation. Only one calibration method can be selected for each method,
which is then valid for all available VA measuring commands.

Tabs

■

General

■

Calibration curves

5.6.5.2

Calibration - General

Tab: Method

▶ Evaluation - Calibration ▶ General

This tab contains the basic settings for the calibration.

Calibration method

Selection of the method with which calibration is to take place.

Selection

no calibration | MLAT | DT

Default value

no calibration

MLAT
MLAT (Modified Linear Approximation Technique) is the default calibra-
tion method for the determination of brightener in electroplating
baths. Only the Linear regression curve type is permitted for this cali-
bration method (see Glossary, page 961).
DT
DT (Dilution Titration) is a calibration method for the determination of
suppressor in electroplating baths. The analysis is based on the fact
that galvanic metal precipitation is reduced by the addition of suppres-
sor to a VMS (Virgin Make-up Solution) (see Glossary, page 958).

5.6.5.3

Calibration - Calibration curves

Tab: Method

▶ Evaluation - Calibration ▶ Calibration curves

Selection of the measuring command

Measuring command

Selection list with all of the VA measuring commands with data acquisi-
tion defined in the method.

Selection

'VA measuring command'