Mini-Optimization Protocols
The following pre-set protocols use four or six well sets in the electroporation plate.
Opt mini 96 well/ Sqr.Exp
Table 7 shows the parameters that vary in each well set in a 96-well plate. The protocol
includes the following well sets:
•
Well set ABCD 1-3: Square wave, 200 V, 2,000 μF, 1,000
Ω, and 20 ms
•
Well set ABCD 4-6: Exponential wave, 250 V, 350–750 μF, and 1000
Ω
Table 7. Opt. mini 96/ Sqr.Exp conditions that vary across the plate.
Square wave
Exponential wave
1
2
3
4
5
6
7
8
9
10 11 12
A
200 V
2,000 μF
20 ms
250 V
2,000 μF
20 ms
300 V
2,000 μF
20 ms
250 V
350 μF
1000
Ω
250 V
500 μF
1000
Ω
250 V
750 μF
1000
Ω
B
C
D
E
F
G
H