Gene Pulser MXcell™ System Manual | Operating Conditions

iv

Mechanical Hazards

The Gene Pulser MXcell system contains a patented arc-protection circuit that dramatically reduces the
incidence of arcing in the cuvette when high voltage is delivered into the sample. The unit incorporates a
circuit that senses the beginning of an arc and diverts current from the sample within <2 μs, preventing or
greatly reducing mechanical, visual, and auditory phenomena at the plate chamber. Should an arc occur,
the sample chamber is effective in containing these small discharges. If you prefer you can wear safety
glasses as additional protection when using the instrument.

Other Safety Precautions

Avoid spilling any liquids onto the apparatus. Use only a paper towel or a cloth wetted with either water or
alcohol to clean the outside surfaces of the Gene Pulser MXcell electroporation system.

Use only the Bio-Rad cables supplied with the Gene Pulser MXcell electroporation system

Use the Gene Pulser MXcell plate chamber only in the assembled condition. Do not attempt to
circumvent the protection of the plate chamber or use it while disassembled.

Read the instruction manual before using the Gene Pulser MXcell electroporation system. For technical
assistance contact your local Bio-Rad office, or in the US call 1-800-4BIORAD (1-800-424-6723).

WARNING! The Gene Pulser MXcell electroporation system generates, uses, and radiates
radio frequency energy. If it is not used in accordance with the instructions given in this
manual, it may cause interference with radio communications. The Gene Pulser MXcell has
been tested and found to comply with the limits for Class A computing devices (pursuant to
Subpart J of Part 15 of FCC Rules) which provide reasonable protection against such
interference when operated in a commercial environment. Operation of this equipment in a
residential area is likely to cause interference. In this case the user will be required, at his/her
expense, to take whatever measure may be required to correct the interference.

Operating Conditions

To safely operate the Gene Pulser MXcell electroporation system, keep environmental conditions within
the following limits:

Operate between 18

o

C and 35

o

C at 90% maximum humidity

Store between -40

o

C and +65

o

C

Mains voltage is 90-132 VRMS and 198-264 VRMS at 47-63 Hz; auto-select

Maximum Input Power is 600 VA