Well set protocols, 96 well/ exp,vgrad/ cgrad, 96 well/ sqr,vgrad, dgrad – Bio-Rad Gene Pulser MXcell™ Electroporation System User Manual

Page 46

Advertising
background image

Well

Set

Protocols

37

Well Set Protocols

Well set protocols include parameters that vary in each well of a 96-well plate. The protocol
includes the following well sets.

96 well/ Exp,Vgrad/ Cgrad

Table 10 shows the parameters that vary in each well of a 96-well plate. The protocol
includes the following well sets:

Well sets ABCD 1–12: Exponential wave, voltage gradient (

ΔV), 350 μF, 1000 Ω

Well sets EFGH 1-12: Exponential wave, current gradient (

ΔC), 250 V, 1000 Ω

96 well/ Sqr,Vgrad, Dgrad

Table 10 shows the parameters that vary in each well set of a 96-well plate. The protocol
includes the following well sets:

Well sets ABCD 1–12: Square wave, voltage gradient (

ΔV), 2000 μF, 1000 Ω, 20 ms

Well sets EFGH 1-12: Square wave, pulse duration gradient (

ΔD) , 250 V, 2000 μF,

1000

Ω

Table 11. 96well/ Exp, Vgrad/Cgrad conditions that vary across the plate.

1

2

3

4

5

6

7

8

9

10

11

12

A

ΔV
(V)

100

100

100

200

200

200

300

300

300

400

400

400

B

C

D

E

ΔC
(μF)

200

200

200

350

350

350

500

500

500

1000

1000

1000

F

G

H

Table 12. 96 well/ Sqr,VgradDgrad conditions that vary across the plate.

1

2

3

4

5

6

7

8

9

10

11

12

A

ΔV
(V)

100

100

100

200

200

200

300

300

300

400

400

400

B

C

D

E

ΔD
(ms)

10

10

10

15

15

15

20

20

20

30

30

30

F

G

H