Opt mini 96 well/ sqr, Opt mini 96 well/ exp – Bio-Rad Gene Pulser MXcell™ Electroporation System User Manual
Page 44
Mini-Optimization
Protocols
35
Opt mini 96 well/ Sqr
Table 8 shows the parameters that vary in each well of a 96-well plate. The protocol
includes the following well sets:
•
Well set ABCD 1-3: Square wave, 200–300 V, 2000 μF, 1000
Ω, 20 ms
•
Well set ABCD 4-6: Square wave, 250V, 2000 μF, 1000
Ω, 15-25 ms
Opt mini 96 well/ Exp
Table 9 shows the parameters that vary in each well of a 96-well plate. The protocol
includes the following well sets:
•
Well set 1-3: 200–300 V, 350 μF, and 1000
Ω
•
Well set 4-6: 250V, 200–500 μF, 1000
Ω
Table 8. Opt mini 96well/sqr conditions that vary across the plate.
Square wave
1
2
3
4
5
6
7
8
9
10
11
12
A
200 V
20 ms
250 V
20 ms
300 V
20 ms
250 V
15 ms
250 V
20 ms
250 V
25 ms
B
C
D
E
F
G
H