Mini-Optimization

Protocols

35

Opt mini 96 well/ Sqr

Table 8 shows the parameters that vary in each well of a 96-well plate. The protocol
includes the following well sets:

Well set ABCD 1-3: Square wave, 200–300 V, 2000 μF, 1000

Ω, 20 ms

Well set ABCD 4-6: Square wave, 250V, 2000 μF, 1000

Ω, 15-25 ms

Opt mini 96 well/ Exp

Table 9 shows the parameters that vary in each well of a 96-well plate. The protocol
includes the following well sets:

Well set 1-3: 200–300 V, 350 μF, and 1000

Ω

Well set 4-6: 250V, 200–500 μF, 1000

Ω

Table 8. Opt mini 96well/sqr conditions that vary across the plate.

Square wave

1

2

3

4

5

6

7

8

9

10

11

12

A

200 V
20 ms

250 V
20 ms

300 V
20 ms

250 V
15 ms

250 V
20 ms

250 V
25 ms

B

C

D

E

F

G

H

Table 9. Opt mini 96 well/exp conditions that vary across the plate.

1

2

3

4

5

6

7

8

9

10

11

12

A

200 V
350 μF

250 V
350 μF

300 V
350 μF

250 V
200 μF

250 V
350 μF

250 V
500 μF

B

C

D

E

F

G

H