3.3 Explanation of Safety Features and Setup Options

3.3.1 Pfd: Power Failure Detection

Preference

Options

Description

PFd

OFF

OFF: Turns power failure detection OFF.

Power Failure Detection

(Default)

In this mode, the run is terminated if a

NEXT RUN

power failure occurs.

ON

NEXT RUN: Turns power failure
detection on for a single run. PFd will
show on the upper right corner of the
Run screen.

ON: Turns power failure detection mode
on in all runs. If there is a power failure in
this mode, then the run will resume when
the power is restored. PFd will show on
the upper right corner of the Run screen.

Warning: Turning the power supply off
(Stopping a run in progress with PFd
enabled, is regarded as a power failure.)
The "interrupted" run will resume
automatically when the power is
restored.

1.

If a power failure occurred during the run with
PFd enabled, the MORE soft key is displayed
on the Run Completed screen.

Press MORE to display details about the most
recent power outage and when power was
restored.

Press RERUN to restart the run.

Press BACK to return to the Run Completed
screen. Press EXIT to go to the Method screen.

2.

Displayed if PFd = OFF

This error screen is displayed if a run has
terminated due to a power failure.

Press RESET to return to the Home screen.

7

SETUP

* PFd

OFF

RRCd

ON

NLd

ON

CHANGE

EXIT

RESET