Error Message

Possible Cause

Solution

ERROR STOP: CODE 09

1. Loose output

1. Verify all electrical

RAPID CHANGE IN R

connections leading

connections.

CORRECT AND CONTINUE

to intermittent

RUN PAUSED

connection to the
loads.

2. Cells added or

2. Pause power supply prior

removed during

to adding or removing

the run.

electrophoresis cells

3. Change in buffer

3. Verify buffer levels are

levels

appropriate.

Note: Certain applications
exhibit intrinsic fluctuations in
resistance (e.g., use of DCode
cell). If this is the case, the load
detection feature can be disabled
to allow uninterrupted completion
of the run. See SETUP, Section 3.

ERROR STOP: CODE 11

Insulation failure in the

Current flow is not desirable

GROUND LEAK

electrical connections

and could be harmful.

CORRECT AND CONTINUE

outside the power supply.

Check electrical connections,

RUN PAUSED

electrophoresis cell, and
chiller system for leaks.
Verify that the electrophoresis
cell rests on an insulated
and dry surface.

Check power source
connections. Additional
capacitance from external EM
filters or an uninterruptible
power supply can cause
excessive ground currents.

ERROR STOP: CODE 12

Improper load or

Press Reset and restart the

REGULATION ERROR

connection to a

run.

RUN TERMINATED

voltage source

If the problem continues,

RESET WHEN READY

cycle power and restart the
run. If the problem persists,
contact Bio-Rad Technical
Service.

ERROR STOP: CODE 13

Internal circuitry

Turn power supply off and on then

PERM MEM ERROR

problems

restart run. If problem persists,
contact Bio-Rad Technical
Support.

ERROR STOP: CODE 16

The run is terminated, either:

Press the RESET soft key and

TPROBE PLUGIN

1. The temperature probe is

restart the run or program a new

ERROR

disconnected from the

run as needed.

RUN TERMINATED

PowerPac during a

RESET WHEN READY

temperature run.

2. The temperature probe

is connected to the
PowerPac during a
non-temperature mode
run.