Bio-Rad PowerPac™ HV High-Voltage Power Supply User Manual

Page 17

Advertising
background image

KEY CHIRP

ON

Activates the audible signal for keystroke.

(Default)
OFF

DATE

MM/DD/YY

This is the only format available.
Press the change soft key then enter the
correct date using the alphanumeric
keypad.

TIME

HH:mm

24 hr format. Factory setting is Pacific
Standard time

Press the change soft key then enter the
correct time using the alphanumeric
keypad.

TEMP PROBE CK

ON

This preference is only available when

Temperature Probe Check

(Default)

the temperature probe is connected.

OFF

ON: The run is paused and error code
18 is displayed in case of failure. A failure
occurs when the temperature remains
5°C less than the set temperature for
more than 15 minutes. This can be caused
by a detached temperature probe or an
insufficient power setting.

OFF: Run will continue regardless the
temperature.

9

SETUP

KEY CHIRP

ON

DATE

12/07/04

* TIME

16:38 HRS

CHANGE

EXIT

RESET


IRDA ON
CONTRAST HI
* KEY CHIRP ON
CHANGE EXIT RESET

SETUP

KEY CHIRP

ON

DATE

12/07/04

* TIME

16:38 HRS

CHANGE

EXIT

RESET

PowerPac HV – Instruction Manual