Bio-Rad PowerPac™ HV High-Voltage Power Supply User Manual

Page 16

Advertising
background image

3.3.2 RRCD: Rapid Resistance Change Detection

Preference

Options

Description

RRCd

ON

ON: Detection of a sudden resistance

Rapid Resistance Change

(Default)

change greater than 20% will cause an

detection

OFF

alarm to sound, the run to pause, and
error message code 09 to be displayed.
(See Section 6.3)

OFF: All resistance changes are ignored
and the run does not stop. When running
in this mode, the

symbol is displayed

on the run
screen as a reminder that this safety
feature has been turned off.
Note: Completion of certain
electrophoresis applications using cells,
such as the DCode

and Rotofor

®

, may

require RRCd to be turned off.

3.3.3 Nld: No Load Detection
NLd

ON

ON: A drop in the current below a set

No Load Detection

(Default)

threshold will cause the run to pause,

OFF

an alarm to sound and the error code 01
to be displayed (see Section 6.3). For
example, if an electrophoresis cell is not
connected or the current is too low (less
than 0.02mA for Manual or Basic
Methods and 0.005mA for IEF methods).

OFF: All load changes are ignored and
the run does not stop. When running in
this mode, the

symbol is displayed

on the Run
screen as a reminder that this safety
feature has been turned off. Some IEF
applications may run at a
current lower than 0.005 mA, therefore,
the No Load detection may need to be
disabled in order to complete the run.

IRDA Port

ON

ON: Enables IR port communication.

Infrared (IR) port

(Default)

communication

OFF

Display CONTRAST

HI

Adjusts the display contrast to

(Default)

accommodate different illumination

ME

conditions.

LO

HI

=

High contrast

ME

=

Medium contrast

LO =

Low

contrast

8

OFF: Disables IR port communication to
prevent communication interference with
other nearby IR devices, e.g., another
PowerPac power supply

PowerPac HV – Instruction Manual