Step Procedure

Description

8.

Select the time mode

Use the soft keys to select a time mode and the
alphanumeric keypad to enter the run time.

HRS:

Run time is in units of hours.

VHOURS:

Run time is in units of volt-hours.

UNTIMED: Run is continuous.

Note: Steps programmed after an untimed step are
ignored.

Press EDIT to return to the Method Screen.

9.

Add a new step or

To program additional steps press the down arrow

save the method

key to select the empty step and repeat Steps 5–8.
If the method will not be saved, go directly to Step 10.

Press SAVE and then go to Step 10.

Press DELETE to delete a selected step.

Press the METHOD soft key to return to the Method
List.

10.

Start the run

Press

to start a run.

11.

Monitor the run

The RUN screen is used to monitor run parameters
or to pause a run. The current step, Volts, mA and
Watts values, elapsed time versus the programmed
time, and total elapsed time are displayed on the
RUN screen. The method step is also indicated.
This run is executing Step 1 of 2.

The run may be paused by pressing the PAUSE
soft key or the

key.

To edit the run parameters during PAUSE see
Section 4.5. Edits made during the run may be
saved, if the SAVE soft key is pressed at the end of
the run. Otherwise the method will remain as
originally created.

To terminate the run, press the

key.

12.

The Run Completed or Run Terminated screen is
displayed at the end of a run. Press RERUN to
repeat the run or EXIT to display the method step
list.