Bio-Rad PowerPac™ HV High-Voltage Power Supply User Manual

Page 41

Advertising
background image

Error Message

Possible Cause

Solution

ERROR STOP: CODE 18

The set temperature cannot

TEMP REG ERROR

be reached because:

CORRECT AND CONTINUE

1. The temperature probe is

1. Make sure the temperature

RUN PAUSED

no longer connected to

probe is securely fastened to

the electrophoresis cell

the electrophoresis cell.

or the temperature probe
made poor contact with
the electrophoresis cell.

2. The programmed power is

2. Adjust the power so it is

insufficient to reach the set

sufficient to reach the set

temperature.

temperature.

Press CONTINUE after the
appropriate corrections are made.

Press CANCEL to terminate the
run

ERROR CODES: 14, 15, 17,

Hardware

Turn power supply off and on then

19–99

malfunction

restart run. If problem persists,
contact Bio-Rad Technical
Support.

Contacting Technical Support

For Technical Support Contact your local Bio-Rad office. In the US call 1-800-4BIORAD
(1-800-424-6723)

Please have the following information available:

Model number: Located on the sticker on the bottom of the unit.

Serial number: Located on the sticker on the bottom of the unit.

Firmware version: The PowerPac HV displays the software version momentarily after
switching the power ON.

State clearly the error code, error message or anomaly, and the conditions that originated
the problem, including run parameters (V, A and W) as well as electrophoresis cell and
buffer system.

Calibration and Service agreements are available through Technical Support.