50 °c, 55 w, 216 v – Bio-Rad PowerPac™ HV High-Voltage Power Supply User Manual

Page 30: 4 temperature mode, 22 powerpac hv – instruction manual

Advertising
background image

Step Procedure

Description

13.

The Run Completed or Run Terminated screen is
displayed at the end of a run. Press RERUN to
repeat the run or EXIT to display the method step
list.

4.4 Temperature Mode

The Temperature operation mode is used to run an electrophoresis apparatus at constant
temperature. Temperature control is very useful to perform DNA sequencing and Single
Stranded Conformational Polymorphism (SSCP) analysis in the Sequi-Gen

®

GT system. The

temperature limit may be set from 30 to 90°C; actual temperature will be controlled from room
temperature to 90°C.

The PowerPac HV monitors the temperature of the electrophoresis cell through the
temperature probe (catalog #165-5058) attached to the wall of the cell. The temperature of
the cell is controlled by regulating the power output (in W) once the cell temperature is
within 2°C of the set temperature.

Fig. 10. Menu Overview for Temperature Mode

TEMPERATURE MODE

TO EXIT TEMP MODE

REMOVE THE PROBE

CONTINUE

SET TEMPERATURE LIMIT

50 °C

______________________

EDIT

EDIT TIME

55 W

5000V

500 mA

TIME: UNTIMED 50°C _

HRS

VHRS UNTIMED

EDIT PARAMETERS

55 W

5000 V

500 mA

TIME: UNTIMED 50°C _

LIMITV LIMITA CONSTW

RUN

216 V

235 mA

50 W

00:02/00:30 01:32HRS_____

PAUSE

EDIT

EDIT

Connect Temperature

Probe to rear of

instrument

22

PowerPac HV – Instruction Manual