Bio-Rad PowerPac™ HV High-Voltage Power Supply User Manual

Page 28

Advertising
background image

Step Procedure

Description

4.

Method name

Name a method.

• Press the up arrow key to select the method

name field.

• Use the CE button to clear the name field and the

alphanumeric keypad to enter a method name.
Refer to Section 2 for help using the keypad.

• Press OK to accept the new name and return to

the METHOD SCREEN with new or edited run
name.

Press CANCEL to abort naming the method and
return to the METHOD SCREEN.

5.

Method steps

Enter the Step Edit sequence.

• Press the up and down arrows to select an

existing step or select the (EMPTY) step to add a
new step.

• Press EDIT to enter the edit loop for the selected

step.

6.

Set the Voltage profile

• Use the V-MAX soft key to enter the maximum

voltage value with the keypad for the current step

• Use the V-SLOPE soft key to select the voltage

ramping (slow, linear or rapid).

Press EDIT to proceed to step 8, Edit Limits.

Use the DELAY soft key, available only in the
first step of a method, to program a run delay. No
voltage is applied during the duration of the step.

7.

Set time delay

• Enter the time delay and press OK to accept, and

go to the method steps screen (step 5) to edit
additional steps.

CANCEL to return to the Volt profile screen.

20

METHOD: pH-3-10

S1 500V

00:05 HRS

S2 5000V

05:00 HRS

*S3 500V

00:30 HRS

METHODS DELETE

S1/1 VOLT PROFILE

V-SLOPE
RAPID

TIME: 01:00 HRS
V-MAX V-SLOPE DELAY

S1/1 EDIT TIME DELAY

DELAY: 00:00 HRS
CANCEL OK

PowerPac HV – Instruction Manual