Step Procedure

Description

5.

Method steps

Press EDIT to enter the edit loop.

New method

• Press the up and down arrows to select an

existing step or select the (EMPTY) step to add a
new step

• Press EDIT again to enter the edit loop for the

selected step

Existing method

6.

Set the constant

Use the soft keys to select the constant parameter

parameter for the

and the keypad to enter its value. Note that the

current step

non-constant values are listed on the right side of
the screen and the time mode at the bottom.

The S#/# indicates which step is being modified. In
this case, S1/1 means editing step 1 of 1.

CONSTV:

Operates with constant voltage.

CONSTA:

Operates with constant current.

CONSTW: Operates with constant power.

Press EDIT to continue.

7.

Enter limits for the

Use the soft keys to select a parameter and the

non-constant

alphanumerical keypad to enter its value.

parameters for the
current step

The soft keys CONST_ or LIMIT_ change according

to the constant parameter selected in the EDIT
CONSTANT screen.

Note: The constant parameter value can be
changed at this point. Press its soft key and enter a
new value.

LIMITV

Change the voltage limit (or constant)

(CONSTV): value.

LIMITA

Change the current limit (or constant)

(CONSTA): value.

LIMITW

Change the power limit (or constant)

(CONSTW): value.

Press EDIT to continue.

16

METHOD: NAME

S1 (EMPTY)

*

METHODS

METHOD: NAME

S1 100V 00:30 HRS

S2 (EMPTY)

*

METHODS

DELETE

SAVE

500 mA
400 W

S1/1 EDIT LIMITS

TIME: UNTIMED
CONSTV LIMIT

A

LIMITW

PowerPac HV – Instruction Manual