4.5.1 Editing a Paused Run
Step Procedure
Description
1.
Pause the run
Press the PAUSE soft key or
to pause a run
that is in progress. The S#/#
indicates this run is
performing Step 1 of 2 steps.
2.
Edit the method
The pause window displays the current run
parameters as well as the date and time that the run
was paused.
Press the EDIT soft key or front panel EDIT button
to edit the run parameters. Follow the instructions in
the appropriate section.
• Section 4.1 for Manual run editing
• Section 4.2 for Basic Methods editing
• Section 4.3 for IEF Methods editing
• Section 4.4 for Temperature mode run editing
Use the CONTINUE soft key or
to resume the
run.
3.
Exit the Run Completed
After the run is finished, press EXIT to return to the
screen
method screen to save the edits (step 4).
Press RERUN to restart the run, keeping the run
parameters the same as the last edit.
4.
Save the changes
From the Method screen, press SAVE to save any
changes made during the run.
Note: Pressing SAVE will overwrite the original
method.
26
Time: 00:00 Vh: 00000
RUN PAUSED AT:
05/27/03 12:29:00
100V 53mA 5W
CONTINUE
EDIT
PowerPac HV – Instruction Manual