4.5 Edit a Paused Run

The PowerPac HV permits editing the run parameters during a pause in all operation
methods modes: Manual, Basic, IEF Methods, and Temperature Mode. The edit capability
during Pause is limited to the current and subsequent steps, and follows the same edit
sequence used when programming the original run. If the current step started as untimed it
cannot be changed to hours or volt-hours.

Fig. 11. Menu Overview for Editing a Paused Method.


RUN: S2/4

216 V

235

mA

50 W

00:02/00:30 :32HRS____

PAUSE

RUN PAUSED AT:
02/18/03 16:11:09
216V 235mA 50W
Time: 00:01 Vh: 00011___
EDIT CONTINUE

RUN TERMINATED AT:
02/18/03

16:11:09

216V 235mA 50W
TIME: 1:00 Vh: 00175
RERUN EXIT

EDIT Loop -
See specific
method section

RUN COMPLETED AT:
02/18/03

16:11:09

216V 235mA 50W
TIME: 1:00 Vh: 00175
RERUN EXIT