6.1.2 Replacement Parts

Catalog
Number

Description

900-8933

Replacement Fuse, 8 A, 250 V, 5 x 20 mm, type T

165-9005

Power Cable

165-5058

Temperature Probe

6.2 Troubleshooting

Problem Possible

Cause

Solution

No display/lights/fan

1. No AC power.

1. Check if the PowerPac HV

is unplugged, or if there is a
problem with the AC power
source, or if the power
switch is in the “off” position.

2. Blown fuse

2. Replace the fuse. See

Section 6.1.1 for details.

Repeatedly blown fuses

Hardware failure

Contact Bio-Rad Technical
Support.

Leads from the cell are

Some leads are not

The PowerPac HV not

not long enough to fit

long enough to make

compatible with any adaptors or

the output jacks

electrical connection.

banana plugs. The PowerPac

The output terminals for the

HV can only be operated safely

PowerPac HV are

with the included connectors.

recessed 27 mm to
meet safety regulations.

Clock settings are lost

Clock battery is low.

Contact Bio-Rad Technical
Support.