4.3.1 Create, Edit, and Run an IEF Method

This section describes how to create and edit methods.

Step Procedure

Description

1.

Set up the power supply

Refer to Section 3 for instrument setup.

2.

Select the operation

At the Home screen, press the IEF METHODS soft

mode

key to display the IEF Method List screen.

3.

IEF Method List

Options:

• Press NEW to create a new method (go to step 4).

New method

OPEN a stored method. Select a stored method

using the arrow scroll keys, an asterisk indicates
the selected method, and press the OPEN soft
key. Go to step 4 to edit method, otherwise go to
step 11 to start the run.

DELETE a stored method. An asterisk indicates

the selected method. Select a stored method
using the arrow scroll keys and press the
DELETE Soft key.

Existing method

Note: If the method list is full (nine methods), a
method must be deleted before a new method
can be added to the list.

19

SELECT MODE

METHODS

MANUAL

IEF

BASIC

IEF METHOD LIST
* pH-3-10

... EMPTY 2 ...
... EMPTY 3 ...

OPEN

DELETE

NEW

METHOD: pH-3-10

S1 500V

00:05 HRS

S2 5000V

05:00 HRS

*S3 500V

00:30 HRS

METHODS DELETE

PowerPac HV – Instruction Manual